IMPLEN NANOPHOTOMETER N50 UV VIS SPECTROPHOTOMETER
En stock

Description
NanoVolume Performance
| Detection Range dsDNA | 1 ng/µl to 16,500 ng/µl (N50: 5 ng/µl to 7,500 ng/µl) |
| Detection Range BSA | 0.03 mg/ml to 478 mg/ml (N50: 0.15 mg/ml to 217 mg/ml) |
| Minimum Sample Size | 0.3 µl |
| Photometric Range (10 mm equivalent) | 0 – 330 A (N50: 0.1 – 150 A) |
| Path Length | 0.67 and 0.07 mm |
| Dilution Factor | 15 and 140 |
| Vortexer | 2,800 rpm; tube size up to 2.0 ml (N50: no vortex included) |
Optical Specifications
| Wavelength Scan Range | 200 – 900 nm (NP50: 200 – 650 nm) |
| Measure Time For Full Scan Range | 3.5 seconds (N50: 6.0 seconds) |
| Wavelength Reproducibility | ± 0.2 nm (N50: ± 1nm) |
| Wavelength Accuracy | ± 0.75 nm (N50: 1.5 nm) |
| Bandwidth | 1.8 nm (N50: 5nm) |
| Stray Light | < 0.5% at 220 nm using NaI |
| Absorbance Reproducibility | < 0.002 A (0.67 mm path) @ 280 nm (N50: 0.004 A (< 0.67 mm path) @ 280 nm) |
| Absorbance Accuracy | ±0.002 A or ±1% of the reading, whichever is greater |
| Zero Stability | ±0.003 A/hour after 20 min warm up @ 280 nm |
| Noise | 0.002 A rms at 0 A @ 280 nm 0.002 A (pk to pk) at 0 A @ 280 nm |
| Optical Arrangement | 1 x 3648 CCD Array |
| Lamp | Xenon flash lamp |
| Lifetime | 109 flashes, up to 10 years |
Compatibility Requirements
| Control Options | Onboard with built-in touch screen, computer, smartphone and tablet |
| Software Compatibility | Windows 7 (32 and 64 bit), Windows 8 (32 and 64 bit), OS X, iOS and Android OS |
| Min. Requirement Smart Phone/Tablet | 4 inch screen; Apple: iPad 2, iPhone5 and iOS 6 Android version 4.4, Quadcore 1.2 GHz with 1 GB RAM |
Documents (1)
-
Paiements sécurisés
Paiement via CB, virement, chèque
-
Livraison
Livraison rapide via TNT
-
Garanties
Garanties légales
-
SAV
Service après-vente basé en France